Recalibrate iNSIGHT 4ES Scantron Scanner

Summary

Calibration automatically adjusts the scanner read system to compensate for variations in scanner components.

Body

When to Use?

Scantron scanners should be recalibrated periodically.  While there is no set standard, we recommend recalibrating weekly.

Procedure

  • Prepare for recalibration
    For best results, perform the following before calibrating:
    • Clean the scanner and read heads.
    • Locate the calibration forms (usually in a white box with instruction sheet)
    • Ensure that the calibration forms are clean and undamaged.
    • Opscan iNSIGHT Calibration Sheets
  • Navigate to the background menu
    In order to recalibrate, begin by navigating to the scanner's background menu.
    Use the following procedure.
     
    1. Enter the background menu by holding down both buttons on the iNSIGHT 4ES Scanner for about five seconds.
      Hold both scantron buttons down to enter background menu
      When the message "Background Menu" appears, release the buttons.
       
    2. Press Next five times or until you reach the Utilities Menu.

      Scantron Utilities Menu
       
    3. Press Enter to enter the Utilities submenu.
       
  • Perform a Top Head Calibrate function
     
    1. Load the calibration forms (part # 202-989-008, pictured in upper right corner)
      by placing them in the input tray and depressing the Press To Load button.
      The orientation of the calibration sheets does not matter.

      Top Head Calibrate
       
    2. Press Run to calibrate the top read head.
       
    3. When calibration is complete, Calibration Complete appears on the screen.

      Calibration Complete
       
    4. Press Next to complete calibration and to advance to the next menu item.
       
  • Perform a Bot Head Calibrate function
    From the Bot Head Calibrate menu item, perform a bottom read head calibration.
     
    1. Press Run to calibrate the bottom read head.

      Bot Head Calibrate
       
    2. When calibration is complete, Calibration Complete appears on the screen.

      Calibration Complete

      NOTE: iNSIGHT 4 ES scanners require 10 calibration sheets for calibration. 
      If you run out of calibration sheets, reload the ones that are in the output tray, and press Start to continue calibration.
      Press Next at any time to abort calibration.
       
  • Exit the background menu
     
    1. Press Next until you reach the Background Menu item.

      Background Menu
       
    2. Press Exit to exit the background menu.

 

Details

Details

Article ID: 16737
Created
Fri 10/13/23 2:04 PM
Modified
Thu 12/26/24 8:01 AM